• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 4.5X36MM PERIPHERAL SCREW, NON-LOCKING; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 4.5X36MM PERIPHERAL SCREW, NON-LOCKING; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number 4.5X36MM PERIPHERAL SCREW, NON-LOCKING
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 08/25/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that (b)(6) 2019 a patient had a reverse shoulder procedure.On (b)(6) 2020 the patient underwent a revision surgery.The patient had an autograft to the glenoid on her initial surgery and it did not appear to have healed.The sales representative who was present during the revision reports that the surgeon had mentioned in surgery that he believed that the locking mechanism on the baseplate may not have been functioning appropriately (the screw was not locked in).During the original procedure the following arthrex devices had been implanted: ar-9564-2436-lat, glenosphere, lot 18.00689.Ar-9503s-03, humeral insert, lot 18.00870.Ar-9502f-36cpc, univers revers suture cup, lot 170141708.Ar-9561-25p, modular central post, lot 5368.Ar-9560-24-2, modular baseplate, lot 5408.Ar-9562-36nl, peripheral screw non-locking, lot 2018001207.Ar-9563-36, peripheral screw locking, lot 2018001490.Ar-9563-40, peripheral screw locking, lot 2018001491.Ar-9501-12p, univers revers humeral stem size 12, lot 170125705.During the revision surgery all off the above listed original implants were explanted, except for the ar-9501-12p, univers revers humeral stem size 12, which was left in the patient.Explanted devices will be returned for evaluation.The surgeon completed the revision by doing a hemi arthroplasty and implanting the following arthrex products: ar-9502-36arca, revers ca adapter assembly, lot 18.01305.Ar-9544-17rca, revers ca humeral head, lot 10071302.Ar-9502f-36cpc, univers revers suture cup, lot 19.00846.The original and revision procedures were performed by the same surgeon at the same facility.
 
Additional Manufacturer Narrative
Complaint not confirmed, despite light damage the device was found to meet dimensional specifications.No abnormality was found that may have contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5X36MM PERIPHERAL SCREW, NON-LOCKING
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key10540478
Report Number1220246-2020-02164
Device Sequence Number16945923
Product Code PHX
UDI-Device Identifier00888867296732
UDI-Public00888867296732
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K173900
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number4.5X36MM PERIPHERAL SCREW, NON-LOCKING
Device Catalogue NumberAR-9562-36NL
Device Lot Number2018001207
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 09/01/2020
Supplement Date Received by Manufacturer09/01/2020
Initial Report FDA Received Date09/16/2020
Supplement Report FDA Received Date09/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-