| Model Number |
24674 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the severely calcified iliac.A 7.0 x 20, 75cm mustang balloon catheter was advanced for dilatation.However, during procedure the balloon ruptured and removed the device without any problem.The procedure was completed with a different device.There were no complications reported and the patient is in good condition.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the severely calcified iliac.A 7.0 x 20, 75cm mustang balloon catheter was advanced for dilatation.However, during procedure the balloon ruptured and removed the device without any problem.The procedure was completed with a different device.There were no complications reported and the patient is in good condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: mustang device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.Blood was identified within the balloon which is evidence of a device leak.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located at the proximal markerband.An examination of the balloon material and proximal markerband identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 20 atmospheres.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no issues or damage to the tip of the device.No other issues were identified during the product analysis.
|
| |
|
Search Alerts/Recalls
|