| Model Number |
M00562531 |
| Medical Device Problem Codes |
Failure to Cut (2587); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The initial reporter address: (b)(6).(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a profile extra small oval flexible snare was used during an unknown procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the snare could be strangulated, but it failed to cut.It seemed that the proximal was bent and felt that tension was applied somewhere.However, it did not transmit to the tip.The procedure was completed with another profile snare.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
|
| |
|
Additional Manufacturer Narrative
|
|
Block e1: the initial reporter address is (b)(6).Block h6: problem code 2587 captures the reportable event of snare loop failure to cut.Block h10: (product investigation): one profle snare was received for analysis.Visual analysis of the device revealed that it did not have any defective condition.Functional evaluation of the returned device found that the device was tested in the 10-inch loop fixture and it was able to extend completely and retract without issues.Continuity test was also performed and the device passed the continuity test since the device's electrical resistance was 8.9 ohms, indicating a proper connection (the electrical resistance shall be less than 20 ohms per non-rotatable snares electrical resistance).A risk review of the proile snare was completed using the snare family global design and confirmed that the event of loop failure to cut and loop bent were defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications, however there is no control how the devices are handled/manipulated in the field.Based on the information available and the analysis performed, the most probable root cause for this problem is no problem detected since the device complaint or problem cannot be confirmed.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a profile extra small oval flexible snare was used during an unknown procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the snare could be strangulated, but it failed to cut.It seemed that the proximal was bent and felt that tension was applied somewhere.However, it did not transmit to the tip.The procedure was completed with another profile snare.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
|
| |
|