• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL FEMORAL NAIL PF, RIGHT T2 ALPHA FEMUR ANTEGRADE Ø12X380MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL FEMORAL NAIL PF, RIGHT T2 ALPHA FEMUR ANTEGRADE Ø12X380MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 2337-1238S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Deformity/ Disfigurement (2360); Impaired Healing (2378)
Date of Event 08/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other additional information is available.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.The device history record could not be reviewed because the affected lot number was not communicated.If any further information is provided, the investigation report will be updated.
 
Event or Problem Description
The manufacturer became aware of a post market clinical follow-up report received from (b)(6), in usa.The title of this report is ¿a retrospective data collection of the treatment of femoral fractures with the femoral nail piriformis fossa (pf) of the t2 alpha femur antegrade gt/pf nailing system¿ which is associated with the stryker ¿t2 alpha femur antegrade gt/pf nailing system¿.This study includes research done on 18 patients (19 cases) requiring surgery between the period march 2019 and october 2019.It was not possible to ascertain specific device details from the report, or to match the events reported with previously reported complaints.Therefore, new complaint was initiated in the system for the post-operative complication mentioned in the report.This product inquiry addresses delayed union.The report states: ¿the 3rd delayed union occurred 152 days after the initial surgery.The patient was a (b)(6) years old female, bmi 24.1, with a history of diabetes and osteoporosis.Her injury was due to standing fall and was classified as 32-a2, closed type.The device used for the surgery was 12x380 t2 alpha femoral nail pf, right with proximal static transverse locking mode, the adverse event is ongoing, and osteogenic stimulator was prescribed.¿.
 
Additional Manufacturer Narrative
Correction: section h6 (health impact code).
 
Event or Problem Description
The manufacturer became aware of a post market clinical follow-up report received from university of louisville, in usa.The title of this report is ¿a retrospective data collection of the treatment of femoral fractures with the femoral nail piriformis fossa (pf) of the t2 alpha femur antegrade gt/pf nailing system¿ which is associated with the stryker ¿t2 alpha femur antegrade gt/pf nailing system¿.This study includes research done on 18 patients (19 cases) requiring surgery between the period (b)(6) 2019.It was not possible to ascertain specific device details from the report, or to match the events reported with previously reported complaints.Therefore, new complaint was initiated in the system for the post-operative complication mentioned in the report.This product inquiry addresses delayed union.The report states: ¿the 3rd delayed union occurred 152 days after the initial surgery.The patient was a 71 years old female, bmi 24.1, with a history of diabetes and osteoporosis.Her injury was due to standing fall and was classified as 32-a2, closed type.The device used for the surgery was 12x380 t2 alpha femoral nail pf, right with proximal static transverse locking mode, the adverse event is ongoing, and osteogenic stimulator was prescribed.¿.
 
Additional Manufacturer Narrative
This complaint has been generated based on findings discovered during post market surveillance literature review.The alleged delayed union with varus at fracture site mentioned in the article could not be confirmed, since no additional information was received from the author or the article.More detailed information about the patient medical history, the event circumstances, and medical reports must be available to determine the root cause.If any additional information becomes available, the investigation will be reopened and re-evaluated accordingly.
 
Event or Problem Description
The manufacturer became aware of a post market clinical follow-up report received from university of (b)(6), in usa.The title of this report is ¿a retrospective data collection of the treatment of femoral fractures with the femoral nail piriformis fossa (pf) of the t2 alpha femur antegrade gt/pf nailing system¿ which is associated with the stryker ¿t2 alpha femur antegrade gt/pf nailing system¿.This study includes research done on 18 patients (19 cases) requiring surgery between the period (b)(6)2019 and (b)(6) 2019.It was not possible to ascertain specific device details from the report, or to match the events reported with previously reported complaints.Therefore, new complaint was initiated in the system for the post-operative complication mentioned in the report.This product inquiry addresses delayed union.The report states: ¿the 3rd delayed union occurred 152 days after the initial surgery.The patient was a 71 years old female, bmi 24.1, with a history of diabetes and osteoporosis.Her injury was due to standing fall and was classified as 32-a2, closed type.The device used for the surgery was 12x380 t2 alpha femoral nail pf, right with proximal static transverse locking mode, the adverse event is ongoing, and osteogenic stimulator was prescribed.¿ additional information received: ¿patient c: delayed union: the 3rd delayed union [.], the adverse event is ongoing, and osteogenic stimulator was prescribed.The patient¿s last follow up was 427 days post op.The bone was healing with some varus at fracture site.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL NAIL PF, RIGHT T2 ALPHA FEMUR ANTEGRADE Ø12X380MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
sharon rivas
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key10541345
Report Number0009610622-2020-00523
Device Sequence Number2962003
Product Code HSB
UDI-Device Identifier07613327289329
UDI-Public07613327289329
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K191271
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number2337-1238S
Device Catalogue Number23371238S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/21/2020
Supplement Date Received by Manufacturer09/30/2020
01/18/2022
Initial Report FDA Received Date09/16/2020
Supplement Report FDA Received Date10/27/2020
02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-