MAUDE Adverse Event Report: MEDTRONIC, INC. MYCARELINK PATIENT MONITOR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. MYCARELINK PATIENT MONITOR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number 24952A
Medical Device Problem Codes Overheating of Device (1437); Output Problem (3005)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 08/31/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the remote monitor was hot and burned the patient's skin when the reader was picked up off the remote monitor.No troubleshooting was specified.The remote monitor is expected to be replaced.No patient complications have been reported as a result of this event.
 
Event or Problem Description
The remote monitor was replaced and returned.
 
Additional Manufacturer Narrative
Product analysis: model#: 24952a, serial/lot#: (b)(6): the remote monitor was returned and analysis revealed a defective radio frequency (rf) head battery during functional test.In addition, the rf head was returned with gold contact.Lastly, the unit was charged for 1 hour and error code 3248 was found in the failure analysis log.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYCARELINK PATIENT MONITOR
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key10541357
Report Number3004593495-2020-01066
Device Sequence Number3110110
Product Code NVZ
UDI-Device Identifier00763000014384
UDI-Public00763000014384
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number24952A
Device Catalogue Number24952A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2020
Supplement Date Received by Manufacturer12/02/2020
Initial Report FDA Received Date09/16/2020
Supplement Report FDA Received Date12/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
-
-