U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AZIYO BIOLOGICS INC. CANGAROO ENVELOPE; MESH, SURGICAL - FTM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AZIYO BIOLOGICS INC. CANGAROO ENVELOPE; MESH, SURGICAL - FTM Back to Search Results
Model Number CMCV-009-MED
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Unspecified Infection (1930)
Date of Event 07/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing review of the cangaroo envelope device history record for the reported lot reflects that all packaged and labeled units were quality released to distribution on 2/12/2020 having met all internal qc acceptance requirements.All sterilization processing records and bioburden testing results indicate a successful sterilization cycle, and passing results of lal and sterility bis allowed the subassembly lot to be released for further packaging and labeling activities.There were no non-conformances during manufacturing or sterilization potentially impacting the final acceptance of this manufacturing lot.In lieu of a request for the oem supplier dhr review, it is noted that aziyo processes the non-sterile envelope materials including cutting, suturing, packaging and sterilizing the product.It is also noted that per the instructions for use (art-20662) provided with the packaged cangaroo envelope device, that infection is listed as a potential complication associated with the surgical procedure and usage of the cangaroo envelope.Although the exact cause of the reported infection cannot be conclusively determined, infection is a known complication associated with the surgical implantation of the cied and the cangaroo envelope devices.Should aziyo receive any additional details related to this event, a supplemental report will be filed.
 
Event or Problem Description
It was reported by aziyo biologics' business partner boston scientific, that a cangaroo envelope (model cmcv-009-med lot# m20b1068) was used with a boston scientific pacemaker (model #/serial # unknown) and implanted on (b)(6) 2020.It was reported that the patient was admitted on (b)(6) 2020 due to infected pacemaker with wound dehiscence.The entire system, including the aziyo cangaroo envelope were explanted on 7/1/2020.Wound was irrigated with antibiotic solution after explanting the device.The issue was reported to aziyo biologics on 8/18/2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANGAROO ENVELOPE
Common Device Name
MESH, SURGICAL - FTM
Manufacturer (Section D)
AZIYO BIOLOGICS INC.
1100 old ellis road
ste 1200
roswell GA 30076
MDR Report Key10541396
Report Number3005619880-2020-00045
Device Sequence Number304175
Product Code FTM
UDI-Device Identifier00859389005034
UDI-Public00859389005034
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K192616
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/06/2022
Device Model NumberCMCV-009-MED
Device Lot NumberM20B1068
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2020
Initial Report FDA Received Date09/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BSC PACEMAKER (MODEL#/SERIAL# - UNKNOWN)
Outcome Attributed to Adverse Event Required Intervention;
-
-