• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE IM ROD; ATTUNE INSTRUMENTS : ALIGNMENT DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE IM ROD; ATTUNE INSTRUMENTS : ALIGNMENT DEVICES Back to Search Results
Model Number 2544-00-502
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907); Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Code Available (3191)
Date of Event 09/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Has product been potentially contaminated or is contaminated with bodily fluids? yes 2545-00-928 has a crack that was discovered during cleaning/steam sterilization.2545-01-969 has broken into 2 pieces that happened per-op.2544-00-502 a piece of plastic fell off during use per-operative.Where / when did the event occur? intra - op, post - op.What action was taken/required to manage the problem during the procedure? with same like product.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned device confirmed the reported event.No evidence was found indicating product malfunction or product error as a contributing factor to the damage and the need for corrective action was not indicated.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE IM ROD
Common Device Name
ATTUNE INSTRUMENTS : ALIGNMENT DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10541471
Report Number1818910-2020-20222
Device Sequence Number18908114
Product Code HWT
UDI-Device Identifier10603295130024
UDI-Public10603295130024
Combination Product (Y/N)N
Initial Reporter CountryNO
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2544-00-502
Device Catalogue Number254400502
Device Lot NumberABG50392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2020
Initial Date Received by Manufacturer 09/02/2020
Supplement Date Received by Manufacturer09/23/2020
10/01/2020
Initial Report FDA Received Date09/16/2020
Supplement Report FDA Received Date09/30/2020
10/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-