RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 16sep2020.
|
|
Event Description
|
The customer reported that the bottom of the touch screen does not respond to touch.The customer contacted product support and was instructed how to do a touchscreen calibration.The customer reported that the unit was not in use on patient.The biomed reported that the issue was discovered while setting the unit for patient use.The biomed stated that there was no delay in therapy.The patient was placed on another vent.The biomed performed touch screen calibration and restored touch, verified correct operation per the touchscreen diagnostics screen.
|
|