MERIT MEDICAL SINGAPORE PTE LTD. MERITRANS DTXPLUS; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE
|
Back to Search Results |
|
Model Number 682000 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
|
Event Description
|
The account alleges that a connection was loose, causing fluid leakage.The connection was re-tightened, but the account reports that it could not be completely fastened, and the leakage continued.No patient injury to report.
|
|
Manufacturer Narrative
|
The suspect device has been returned for evaluation.The device was examined visually, and functional testing was performed no leakage, loose connection or any failure observed on the product returned.The complaint could not be confirmed.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
|
|