MERIT MEDICAL SINGAPORE PTE LTD. MERITRANS DTXPLUS; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE
|
Back to Search Results |
|
Model Number 682000 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/18/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
|
Event Description
|
The account alleges that there are cracks, and leakage near the connection between the sensor and the infusion set.No patient injury to report.
|
|
Manufacturer Narrative
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined however, it is likely that significant force was applied to the device during use.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
|