| Model Number |
0502704030 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a blurry image.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: water intrusion confirmed failure: outer tube damaged (bent, dented),damaged distal cover glass probable root cause: poor mate between cannula and endoscope (i.E.Speedlock or j-lock feature).Poor fit between endoscope needle and cannula.Use error.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a blurry image.
|
| |
|