Model Number BELLAVISTA 1000E
Device Problem Computer Operating System Problem (2898)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/21/2020
Event Type  malfunction  
Manufacturer Narrative
At this time, the suspect device has not been returned for evaluation.However, screen shot of service screen has been sent and analyzed by technical support.It was suspected that the mainboard needs to be replaced but on the other hand they couldn't think of any reason how the settings should have changed no further details received from the user.No root cause has been determined yet because the investigation is still on going.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported blue screen accompanied by a high priority visual/audible alarm with the bellavista 1000e.Ventilation continued, but the mode seemed to have changed from psv (pressure support) to controlled based on the connected patient's chest movement and reaction.The end user restarted the device and went through setup again.The patient was nervous, but unharmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BELLAVISTA
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
IMTMEDICAL AG
gewerbestrasse 8
buchs, 9470
SZ  9470
Manufacturer (Section G)
TECHNOCOM SYSTEMS SDN BHD
plo 1 jaian firma 1
kawasan perindustrian tebrau 1
johor bahru, 81100
MY   81100
Manufacturer Contact
peter sparacio
26125 n. riverwoods blvd.
mettawa, IL 60045
8333273284
MDR Report Key10543819
MDR Text Key207341973
Report Number3004553423-2020-00661
Device Sequence Number1
Product Code CBK
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
K163127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBELLAVISTA 1000E
Device Catalogue Number301.100.100
Initial Date Manufacturer Received 08/21/2020
Initial Date FDA Received09/16/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-