• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R Back to Search Results
Model Number 5DCS
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned to stryker sustainability solutions for evaluation.The device was damaged by the carrier and was unable to be delivered.As the device was not returned for evaluation, inspection was unable to be performed.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The reported event may be attributed to: failure mode: damage to the instrument or the insulation.Use error: user attempts to cut excessive amount of tissue.Damaged insulation due to misuse, including improper insertion to / withdrawal from a trocar.The instructions for use (ifu) state: a complete understanding of the principles and techniques involved in laser, electrosurgical, and ultrasonic procedures is essential to avoid shock and burn hazards to both patient and medical personnel and damage to other medical instruments.Ensure that insulation or grounding is not compromised.Do not immerse electrosurgical instruments in liquid.Do not introduce or withdraw the instrument through a trocar sleeve with the blade/jaws open.The instrument will operate with electrosurgical generators having a high frequency maximum voltage of 3000 volts peak.Ensure that all safety precautions are followed and refer to the electrosurgical generator¿s specification to verify compatibility and for indications and instructions.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
 
Event or Problem Description
It was reported that the insulation broke off from the laparoscopic scissors during procedure.The insulation fell into the patient and was not retrieved.There was no patient injury, medical intervention or extended procedure time reported.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
A review of mdrs filed since may 20, 2019, through september 2, 2022, was conducted.Since august 17, 2020, there have been no additional reports of serious injury or surgical intervention to preclude serious injury or permanent impairment related to this specific failure mode.Based on the analysis of mdr data compared to the firm¿s sales figures, the likelihood of this specific failure mode causing or contributing to another serious injury should the malfunction occur is considered remote.As such, we will no longer be reporting for the failure mode of laparoscopic (lap) broken and detached shaft pin region components.
 
Event or Problem Description
It was reported that the insulation broke off from the laparoscopic scissors during procedure.The insulation fell into the patient and was not retrieved.There was no patient injury, medical intervention or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key10544229
Report Number0001056128-2020-00065
Device Sequence Number3161357
Product Code NUJ
UDI-Device Identifier00885825013288
UDI-Public00885825013288
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K110189
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/11/2023
Device Model Number5DCS
Device Catalogue Number5DCSRR
Device Lot Number11889254
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/18/2020
Supplement Date Received by Manufacturer09/02/2022
Initial Report FDA Received Date09/16/2020
Supplement Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-