Model Number 175782
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Tissue Damage (2104); Blood Loss (2597)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mdr: 9618003-2020-13845 / device 14 of 19.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: (b)(4).
 
Event or Problem Description
The end user estimated about 3 pouches from each of the 6 market units that were affected.The precut opening in the convex mass was very off centered.She stated that the off centered barrier had caused her to develop an open sore at the 9 o'clock position, approximately 13 x 25mm with some bleeding.Her wear time was 3 days.Reportedly, she had been seen by a local ostomy nurse.She was wearing a hernia belt, but recently stopped using the belt.She further reported that the sore now appeared to be healing with new skin and there was no bleeding with last wafer change.The patient continued to use the product.Photographs depicting the issue were received from the end user.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: d4: unique identifier (udi) h6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions h10: investigation summary an investigation was completed for this complaint by the assigned manufacturing site investigation team.The complaint summary of this investigation required rework.A corrective actions/preventive actions (capa) was raised, and an updated/corrected investigation summary has been completed in accordance with assigned capa.A supplemental mdr is being submitted to document the completion of investigation rework and final investigation; the complaint record will proceed to closure.Batch record review: lot 0c02789 was manufactured on 04/14/2020, in the convex 1pc line with a total of (b)(4) ea.Compliance engineer performed a batch record review on 02/26/2021, to verify if all the applicable procedures were followed and no issues were found, all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, under international commodity code (icc) code 175782, system application product (sap) material identification (id) 1000626 and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.Photograph, video and/or physical sample evaluation: there are photographs associated with this case and in these, the reported defect can be seen.No unused return sample was expected.Conclusion summary of the related event: based on the investigation findings, process observation, tooling analysis and personnel interviews, the investigation concluded that the off center on active life products of convex 1pc is originated due to the following root causes: method: - method for wafer placement onto the cups of the lower level of the rotary table is too general.Due to that a wafer placement variation could be found between the different operators, which can potentially cause a wrong placement of wafers, resulting in a nonconforming product impacted by the off-center defect.- method for pouch placement onto slides plates of the upper level of the rotary table is too general.Due to that a pouch placement variation could be found between the different operators, which can potentially cause a wrong placement of wafers resulting in a nonconforming product impacted by the off-center defect.- not standardized method for the installation of slide plates into the rotary table of convex 1pc.- not standardized method for the installation of cups into the rotary table of convex 1pc.Machine: -worn out guides for slide plates, do not allow guides to properly adjust during the manufacturing process, causing misalignment between pouches and wafers.- incorrect adjustment of cups due to the use of the worn-out screws.- rotation of the glue welding heads causes wafers and pouches decentralization during the assembly process.Opportunity of improvement in the detection method: there is not a visual aid for the off-center defect recognition for active life pouches, in the manufacturing line of convex 1pc.A corrective actions/preventive actions (capa) plan was generated for the mitigate the root causes identified.The investigation associated with related event has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVELIFE
Common Device Name
POUCH, COLOSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key10545587
Report Number9618003-2020-13845
Device Sequence Number11547079
Product Code EZQ
UDI-Device Identifier00768455181285
UDI-Public00768455181285
Combination Product (Y/N)N
Initial Reporter StateWI
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/13/2025
Device Model Number175782
Device Lot Number0C02789
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/27/2020
Supplement Date Received by Manufacturer12/01/2023
Initial Report FDA Received Date09/17/2020
Supplement Report FDA Received Date12/04/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
-
-