| Model Number |
87771 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned instrument is currently being evaluated.A follow up report with results of the investigation will be submitted upon completion.
|
| |
|
Event or Problem Description
|
|
The distal tip of an arsenal navigation driver sheared off while attempting to inserter a screw.The detached section was removed but was discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the returned polyaxial screwdriver found that the distal tip had fractured and became separated from the instrument.Inspection under magnification revealed circular fracture lines indicative of a torsional overload.Alphatec spine navigated reusable instruments are intended to be used during the preparation and placement of alphatec screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures.
|
| |
|