SMITH & NEPHEW, INC. ACCORD DUAL ENDED HEX DRIVER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71360046
Device Problem Degraded (1153)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/19/2020
Event Type  malfunction  
Event Description
It was reported that during thr, screwdriver hex heads were stripped making it difficult to use.Delay of less than 30 minutes reported and no patient injuries.Procedure was completed with a s+n back-up device.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damaged and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further information is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCORD DUAL ENDED HEX DRIVER
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10546669
MDR Text Key207324964
Report Number1020279-2020-04637
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010611543
UDI-Public03596010611543
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71360046
Device Catalogue Number71360046
Device Lot Number18JK00001
Was Device Available for Evaluation? No
Date Manufacturer Received05/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-