• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SERVICE REPL CAMERA HEAD LENS INT SYS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SERVICE REPL CAMERA HEAD LENS INT SYS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 72203967S
Device Problem No Display/Image (1183)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed and a damaged edge connector was observed.A functional evaluation revealed that the device switched from the color bars to live video where it displayed a black screen.The complaint has been confirmed and the root cause has been associated with an electrical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that there was fog inside the camera head.It is unknown whether the event happened during surgery and if there was a patient involvement.Results of investigation have concluded that this unit was displaying a black screen which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVICE REPL CAMERA HEAD LENS INT SYS
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10547890
MDR Text Key207364378
Report Number1643264-2020-01198
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00885556631980
UDI-Public00885556631980
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153606
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72203967S
Device Catalogue Number72203967S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/10/2020
Initial Date FDA Received09/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-