| Model Number |
3542914 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
08/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.(b)(4).Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) african american female who underwent primary breast augmentation with a mentor siltex contour profile moderate 425/475 saline prosthesis experienced left sided pain and right sided deflation post procedure.A physical examination was performed by a physician and deflation was diagnosed.As a result patient underwent bilateral replacements with cat#: 3503380 on (b)(6) 2020.This report relates to the right prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on october 8 2020: during the visual evaluation of the device, a tear was observed on the posterior aspect, measuring approximately 1.5 cm.Microscopic examination in the tear edges revealed parallel striations that are consistent with markings made by a sharp instrument perforating silicone material.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.All the implants are 100% inspected before leaving the facility.There is no evidence that the issue is related with manufacturing breast implants are not considered lifetime devices and deflation/rupture is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|