| Model Number |
1012452-12 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was reportedly discarded.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that an nc trek box with the outer labeling of 3.75x12mm contained an nc trek balloon with the inner labeling of a 3x15mm nc trek.The device was not used and there was no patient involvement.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.The reported device markings / labelling problem was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that a mix-up of device packaging occurred at the hospital.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|