DEPUY IRELAND 9616671 METAGLENE GUIDE PIN DIA 2.5MM; EXTREMITY INSTRUMENTS : PIN GUIDES Back to Search Results
Model Number 2307-87-004
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The hole in the metaglen positioner plate (230787003) was defective and the metaglen central guide pin (230787004) could not be inserted through it.For this reason, a 2.0mm wire from the hospital had to be used to continue the operation.
 
Additional Manufacturer Narrative
This product was reported in error.No patient death, serious injury or evidence of reportable product malfunction occurred.Depuy considers the investigation closed at this time.Should any additional information be received to change the outcome of the performed investigation, the complaint will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METAGLENE GUIDE PIN DIA 2.5MM
Common Device Name
EXTREMITY INSTRUMENTS : PIN GUIDES
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
MDR Report Key10549063
Report Number1818910-2020-20405
Device Sequence Number3134346
Product Code LXH
UDI-Device Identifier10603295116455
UDI-Public10603295116455
Combination Product (Y/N)N
Initial Reporter CountrySZ
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2307-87-004
Device Catalogue Number230787004
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/03/2020
Supplement Date Received by Manufacturer09/10/2020
Initial Report FDA Received Date09/17/2020
Supplement Report FDA Received Date09/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-