U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS PER-Q-CATH PICC CATHETER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS PER-Q-CATH PICC CATHETER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problem No Display/Image (1183)
Patient Problem Cardiac Tamponade (2226)
Event Type  Injury  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event Description
It was reported that "today the first day of the training was held and it is possible to identify that they are in serious trouble locating the tip of the catheter in the chest x-ray, due to the visibility of the institution's radiography equipment.They had a serious adverse event, where the insertion of the catheter was performed in the neonatology unit without confirmation of the tip of the catheter, and then was released for use, later the client presented cardiac tamponade, it was verified by contrast x-ray that the catheter was positioned in the right ventricle.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PER-Q-CATH PICC CATHETER
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX  
Manufacturer Contact
kelsey erickson
605 n. 5600 w.
salt lake city, UT 84116
8015225937
MDR Report Key10549667
MDR Text Key207499398
Report Number3006260740-2020-03297
Device Sequence Number1
Product Code LJS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Hospital
Initial Date Manufacturer Received 08/24/2020
Initial Date FDA Received09/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-