WRIGHTS LANE SYNTHES USA PRODUCTS LLC UNK - BONE STAPLE: SPEEDTITAN; STAPLE, FIXATION, BONE
|
Back to Search Results |
|
Device Problem
Break (1069)
|
Patient Problems
Failure of Implant (1924); No Code Available (3191)
|
Event Type
Injury
|
Manufacturer Narrative
|
This report is for an unknown bone staple/ unknown lot.Part and lot number are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/ investigation.The implant(s) was not returned and instead the investigation will be done based on the supplied image(s).The image(s) was reviewed and the complaint condition for broken could be confirmed as the image provided shows a broken staple.As the instrument(s) was not returned an as received condition, dimensional inspection, material or drawing reviews are not applicable.A definitive assignable root cause could not be determined based on the provided information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that on (b)(6) 2020 the patient underwent a revision and a hardware removal due to a broken speedtitan.A speedtitan and three (3) cortex screws were removed and were revised to stryker plates.The procedure was successfully completed.The patient status is unknown.This report is for one (1) unknown bone staple.This is report 1 of 4 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: d7: explantation date updated.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|