| Model Number |
383346 |
| Date of Event |
08/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the bd saf-t-intima¿ iv catheter safety system tubing had a hole in it during use.This complaint was created to capture the 1st of 2 related incidents.The following information was provided by the initial reporter: "tubing attached to y adapter had a hole in it, requiring the iv to be replaced ".
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: a device history record review was completed by our quality engineer team for provided lot number 9326780.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.H3 other text : see h.10.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd saf-t-intima¿ iv catheter safety system tubing had a hole in it during use.This complaint was created to capture the 1st of 2 related incidents.The following information was provided by the initial reporter: "tubing attached to y adapter had a hole in it, requiring the iv to be replaced ".
|
| |
|