• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R Back to Search Results
Model Number LF1744
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is currently in progress.The device has not yet been returned to stryker sustainability solutions for evaluation.A supplemental mdr will be submitted once the investigation is completed.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the customer had several damaged ligasure devices.There were error codes on the generator indicating an incomplete sealing cycle.Multiple attempts were made to obtain additional information from the complainant.The complainant was unable to provide specific event details.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
The investigation results confirm this incident no longer falls within mdr reportability requirements as there were no short circuits identified.The complaint device was subsequently returned to stryker sustainability solutions.Manual continuity tests were performed and confirmed to be acceptable.The 7 spacer pads were inspected for damage and damage was found on 2 spacer pads.A review of the dhr for the reported lot number supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: ancillary equipment failure.Trigger button (activation button) not engaged throughout the entire seal cycle.Device activated in contact with pooling fluid.Device used on vessels thicker than 7 mm.Grasping on too much tissue or inappropriate tissue types or on staples.The instructions for use (ifu) state: use caution during surgical cases in which patients exhibit certain types of vascular pathology (atherosclerosis, aneurysmal vessels, etc.).For best results, apply the seal to unaffected vasculature.If the generator provides multiple power settings, use the lowest power needed to achieve the intended effect.Do not use this instrument on vessels in excess of 7 mm in diameter.If the instrument shaft is visibly bent, discard and replace the instrument.Do not place the vessel and/or tissue in the jaw hinge.Place the vessel and/or tissue in the center of the jaws.Eliminate tension on the tissue while sealing and cutting to ensure proper function.Use caution when grasping, manipulating, sealing, and dividing large tissue bundles.Do not bend instrument shaft.Do not attempt to seal over clips or staples as incomplete seals/damage to the cutting blade will occur.Contact between an active electrode and any metal objects may result in alternate site burns or incomplete seals.Do not overfill the jaws of the instrument with tissue, as this may reduce device performance.Keep the instrument jaws clean.Build-up of eschar may reduce the seal and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.The lever must be continually held with the activation button fully depressed until the seal cycle is complete.The lever does not latch into the activation position.A continuous tone sounds to indicate the activation of rf energy.When the activation cycle is complete, a two-pulsed seal-cycle-complete tone sounds and rf output ceases.Notice ¿ the surgeon may inspect the seal before cutting the vessel or tissue.After inspecting the seal, the surgeon should create a second seal adjacent to the first seal before cutting, as described below.A tone with multiple pulses indicates that the seal cycle was not completed.Refer to the troubleshooting section for possible causes and corrective actions.Do not cut tissue until you have verified that there is an adequate seal.To seal adjacent tissue, overlap the edge of the existing seal.The second seal should be distal to the first seal to increase seal margin.Activating energy delivery with a footswitch when the activation button is not fully depressed may result in improper sealing and increase thermal spread to tissue outside the surgical site.Proper pressure is being applied to the tissue when the lever keeps the activation button fully depressed.Energy-based devices, such as esu pencils or ultrasonic scalpels that are associated with thermal spread should not be used to transect seals.Failure to maintain steady pressure on the lever while cutting can result in inadvertent reactivation of energy.Wipe jaw surfaces and edges with a wet gauze pad as needed.Remove any embedded tissue from blade track and jaw hinge area.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the customer had several damaged ligasure devices.There were error codes on the generator indicating an incomplete sealing cycle.Multiple attempts were made to obtain additional information from the complainant.The complainant was unable to provide specific event details.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
MDR Report Key10551389
Report Number0001056128-2020-00067
Device Sequence Number3345385
Product Code NUJ
UDI-Device Identifier07613327375800
UDI-Public07613327375800
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K180499
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/07/2023
Device Model NumberLF1744
Device Catalogue NumberLF1744RR
Device Lot Number11554819
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2020
Initial Date Received by Manufacturer 08/19/2020
Supplement Date Received by Manufacturer09/30/2020
Initial Report FDA Received Date09/17/2020
Supplement Report FDA Received Date10/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-