| Model Number |
ZCB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Capsular Bag Tear (2639); Vitrectomy (2643)
|
| Date of Event |
08/24/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information: patient age/date of birth: unknown as information was not provided.Patient gender: unknown as information was not provided.If implanted: not applicable, as the intraocular lens was inserted and removed in the initial surgery.If explanted: not applicable, as the intraocular lens was inserted and removed in the initial surgery.Attempts were made to contact the customer account requesting additional information regarding complaint however, to date no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported intraocular lens (iol) was in the injector and in eye but it was decided at last minute to use anterior chamber (ac) lens as the bag was broken and vitrectomy was performed.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received, and the following fields were updated accordingly: section d-10: device available for evaluation: yes.Section d-10: returned to manufacturer on: 09/21/2020.Section h-3: device returned to manufacturer: yes.Device evaluation: the complaint lens was received inside its original lens case.The original folding carton and directions for use (dfu) were received as well.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics, and that the lens was received cut in pieces, which is consistent with a lens that was handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification.Historical data analysis: a search revealed that no other complaints for this production order number have been received.Conclusion: as a result, of the investigation there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|