• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1-PIECE; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1-PIECE; MONOFOCAL IOLS Back to Search Results
Model Number ZCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643)
Date of Event 08/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information: patient age/date of birth: unknown as information was not provided.Patient gender: unknown as information was not provided.If implanted: not applicable, as the intraocular lens was inserted and removed in the initial surgery.If explanted: not applicable, as the intraocular lens was inserted and removed in the initial surgery.Attempts were made to contact the customer account requesting additional information regarding complaint however, to date no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported intraocular lens (iol) was in the injector and in eye but it was decided at last minute to use anterior chamber (ac) lens as the bag was broken and vitrectomy was performed.No further information was provided.
 
Additional Manufacturer Narrative
Additional information was received, and the following fields were updated accordingly: section d-10: device available for evaluation: yes.Section d-10: returned to manufacturer on: 09/21/2020.Section h-3: device returned to manufacturer: yes.Device evaluation: the complaint lens was received inside its original lens case.The original folding carton and directions for use (dfu) were received as well.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics, and that the lens was received cut in pieces, which is consistent with a lens that was handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification.Historical data analysis: a search revealed that no other complaints for this production order number have been received.Conclusion: as a result, of the investigation there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e. st. andrew place
santa ana CA 92705
MDR Report Key10551558
Report Number2648035-2020-00696
Device Sequence Number11365146
Product Code HQL
UDI-Device Identifier05050474531802
UDI-Public(01)05050474531802(17)240603
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/03/2024
Device Model NumberZCB00
Device Catalogue NumberZCB0000245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2020
Initial Date Received by Manufacturer 08/24/2020
Supplement Date Received by Manufacturer09/29/2020
Initial Report FDA Received Date09/17/2020
Supplement Report FDA Received Date10/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-