Model Number H-1129
Medical Device Problem Code Electrical Power Problem (2925)
Health Effect - Clinical Codes No Patient Involvement (2645); No Information (3190)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a smiths medical fluid wramer pulls to much amperage from the putlets and trips the breakers in the shop and the meters in the or no adverse patient effects were reported.
 
Additional Manufacturer Narrative
Additional information- added to event information in section b5.
 
Event or Problem Description
No patient involvement.
 
Additional Manufacturer Narrative
Other, other text: device evaluation: one level 1 normoflo irrigation fast flow system was returned for analysis.Visual inspection performed, and device found with regulator box damages.Visual inspection and testing of all components with multimeter, could not find damages components and all components passed electrical testing.Investigation could not duplicated customer reported reason.Investigator replaced device line cord, main pcb, actuator pole, actuator membrane switch, aux outlet, and rolling base compressor, power supply as preventative action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1
Common Device Name
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
MDR Report Key10552516
Report Number3012307300-2020-09481
Device Sequence Number2803274
Product Code LGZ
UDI-Device Identifier50695085812904
UDI-Public50695085812904
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K072080
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH-1129
Device Catalogue NumberH-1129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/19/2020
Supplement Date Received by Manufacturer09/29/2020
11/18/2020
Initial Report FDA Received Date09/18/2020
Supplement Report FDA Received Date10/28/2020
12/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-