| Model Number |
72200873S |
| Medical Device Problem Code |
Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed and observed no issues.A functional evaluation revealed a hand piece stall error message.The complaint was confirmed and the root cause has been associated with an electrical component failure.Factors that could have contributed to the reported event include connecting a hand piece that is still wet from sterilization processing or connecting a shorted out hand piece causing damage to the main board.The dyonics power ii control system operations/service manual states: to prevent damage to the control unit connector ports, do not plug in wet.Ensure that cleaned or sterilized cable connectors are completely dry prior to connecting to the control unit.
|
| |
|
Event or Problem Description
|
|
It was reported that the dyonics was not working.It is unknown if there was a delay.It is unknown whether the event happened during surgery and if there was patient involvement.No further information.Results of investigation revealed that during functional evaluation the device displayed a handpiece stall error which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|