• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 VIT E NEUTRAL LNR 32MM F; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 VIT E NEUTRAL LNR 32MM F; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Information (3190)
Date of Event 08/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported incoming inspection team member found there is a hole on the outer package and it was penetrated into the sterile package.No additional information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, d10, g4, h2, h3, h6.Corrected: h4.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product confirmed the presence of a hole in the top panel of the carton box.The inward puncture runs through both tyvek lids and the inner cavity.Sterility is compromised.The shrink wrap is missing and the tamper proof label has been cut.Device history record (dhr) was reviewed and no discrepancies were found.The likely condition of the part when it left zimmer biomet control is considered conforming based on the evaluation of the returned product and the dhr review.The root cause of the reported issue is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 VIT E NEUTRAL LNR 32MM F
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10553725
Report Number0001825034-2020-03545
Device Sequence Number3749722
Product Code LPH
UDI-Device Identifier00889024519862
UDI-Public(01)00889024519862(17)250531(10)64749996
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K190660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberN/A
Device Catalogue Number30103206
Device Lot Number64749996
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/26/2020
Supplement Date Received by Manufacturer11/23/2020
Initial Report FDA Received Date09/18/2020
Supplement Report FDA Received Date11/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-