Model Number 284004
Device Problem Decrease in Pressure (1490)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).
 
Event Description
Repair of loan unit, shows "low pressure".The defect was detected at the service center after the loaner has been returned from customer.No patients consequence, no surgical delay.Additional details provided by the affiliate reported the failure was detected during service and not during a procedure.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the device was received and evaluated at the service center.There was no allegation of malfunction against the device from the customer, however, defects were found with the device during service evaluation.Hence this complaint can be confirmed.It was found that the device showed "low pressure" as the solenoid pressure valve was defective.The irreparable valve was replaced and the device was cleaned, repaired, tested and found to be fully functional.However, given the information provided, we cannot determine a definitive root cause for the defective pressure valve.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 12/28/2018 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP-SHAVER BOX
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10554877
MDR Text Key207570522
Report Number1221934-2020-02663
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705027934
UDI-Public10886705027934
Combination Product (y/n)N
PMA/PMN Number
K171237
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Type of Report Initial,Followup
Report Date 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number284004
Device Catalogue Number284004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2020
Initial Date Manufacturer Received 09/02/2020
Initial Date FDA Received09/18/2020
Supplement Dates Manufacturer Received09/30/2020
Supplement Dates FDA Received09/30/2020
Patient Sequence Number1
-
-