LIVANOVA USA, INC. PROGRAMMING SOFTWARE Back to Search Results
Model Number 3000
Medical Device Problem Codes Computer Software Problem (1112); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Voluntary medical device correction (remedial action) was initiated by the manufacturer on 2/4/2020, and the physician was provided a notification letter with recommended actions.Internal reference number: (b)(4).
 
Event or Problem Description
It was reported that the patient had the vns turned off before a mri and the session report showed normal diagnostics, all output currents were off.When the device was turned back on, the device showed a low output current warning, output status was 2.25 ma with a red warning sign, and impedance was ok.The device was later re-interrogated and no errors were present.Internal testing and data review identified that false low output current messages can occur when m3000 (b)(4) software programs a m103-106 generator after a specific programming sequence occurs.When m3000 (b)(4) tablets initiate a programming session with a m103-106 generator, where its output current is programmed off (0 ma), and then the output current is programmed back on to ma, a low output current message would be seen when an in-session interrogation is performed.This would persist upon multiple in-session interrogations and if diagnostics weren't performed during the session, show on the session report.Running system diagnostics or ending and restarting the session will confirm functionality of the device, and resolve the ¿output current low¿ error message.The low output current messages in these cases are false and do not impact generator function.Based on the reported programming sequence that occurred that day, it was consistent with the false low output current mechanism described above.Therefore, the patient's generator is expected to be able to deliver therapy as programmed.No further relevant information has been received to date.
 
Additional Manufacturer Narrative
H7.Corrected information, initial report: inadvertently did not select 'notification'.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Common Device Name
PROGRAMMING SOFTWARE
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
MDR Report Key10554902
Report Number1644487-2020-01244
Device Sequence Number3749793
Product Code LYJ
UDI-Device Identifier05425025750535
UDI-Public05425025750535
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3000
Was Device Available for Evaluation? No
Event Location Other
Initial Date Received by Manufacturer 08/26/2020
Supplement Date Received by Manufacturer10/22/2020
Initial Report FDA Received Date09/18/2020
Supplement Report FDA Received Date10/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age40 YR
-
-