• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 PERFIX PLUG; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 PERFIX PLUG; SURGICAL MESH Back to Search Results
Catalog Number 0112960
Medical Device Problem Codes Defective Device (2588); Insufficient Device Problem Information (3190); Migration (4003)
Health Effect - Clinical Codes Abscess (1690); Hematoma (1884); Unspecified Infection (1930); Injury (2348)
Date of Event 12/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol perfix plug on (b)(6) 2018.As reported, the patient is making a claim for an adverse patient outcome against the perfix plug.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: h11: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-oct-2017) is considered to be a best estimate.Updated fields: a2, b3, b4, b5, b7, d3, d4, d.6b (date of explant) g3, g6, h2, h4, h6, h10, h11 corrected field: d.6a (date of implant) note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol perfix plug on (b)(6) 2018.As reported, the patient is making a claim for an adverse patient outcome against the perfix plug.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2018 - patient was diagnosed with left inguinal hernia thereby underwent open repair with implant of perfix plug and a synthetic mesh.Per op notes, ¿the patulous deep internal ring with incarcerated omentum was identified and reduced.The patulous deep internal ring was reapproximated with perfix plug using sutures.The floor was patched with synthetic mesh and then tacked to pubic tubercle.¿ (b)(6) 2019 - patient was diagnosed with infected mesh and abscess thereby underwent open repair with excision of perfix plug and synthetic mesh.Per op notes, ¿the pus and subcutaneous tissue at the top the prior mesh was taken down.The hematoma was drained.The infected sinus tract was excised.The synthetic mesh was split open, some free-floating mesh with sutures were identified and excised all the meshes.The wound vac was placed.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFIX PLUG
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key10554908
Report Number1213643-2020-08520
Device Sequence Number3220134
Product Code FTL
UDI-Device Identifier00801741016660
UDI-Public(01)00801741016660(17)221128(10)HUBX1575
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K922916
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/28/2022
Device Catalogue Number0112960
Device Lot NumberHUBX1575
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2020
Supplement Date Received by Manufacturer07/16/2025
Initial Report FDA Received Date09/18/2020
Supplement Report FDA Received Date08/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/15/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexMale
-
-