The user facility's biomed department inspected the table and determined the wing nuts for the head section were not properly tightened on the head section during the time of the reported event.The biomed department tightened the wing nuts.A steris service technician arrived onsite to inspect the 4085 surgical table.The technician tested the table and found no issues with the function or operation of the table including the head section.The 4085 surgical table operator manual states (1-2), "warning - personal injury hazard: when installing any table accessory, check for correct attachment and tighten securely (if appropriate).Do not use a worn or damaged accessory.Before using any accessory, ensure it has been installed properly." a steris account manager has offered an in-service training on the proper use and operation of the surgical table, specifically properly attaching the head section; the in-service training is scheduled for (b)(6) 2020.No additional issues were reported.
|
The user facility reported that during a procedure with their 4085 surgical table slid fully to the head end, user facility personnel commanded their table to trendelenburg when the head section slid out of position.As the head section slid out of position, the patient began to slide off the table causing the table to tip downwards.The procedure was cancelled after the reported event.No report of injury.
|