| Model Number |
656416 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "angular stability is functionless.".
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: refer to d4 catalog 3, gtin.The reported event could be confirmed.The inspection revealed a broken/deformed thread winding at the thread outlet, causing that the screw could not be locked into the plate hole.A non-conformity report (nc) was already initiated based on previous complaints reporting the same event.The nc investigation is not finished yet.A review of the device history for the reported lot did not indicate any abnormalities; the screw was manufactured within specifications.A review of the risk management file showed that the reported event is adequately addressed.A review of the labeling did not indicate any abnormalities.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
As reported: "angular stability is functionless.".
|
| |
|
Search Alerts/Recalls
|