• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MS1-4590S
Medical Device Problem Codes Corroded (1131); Mechanical Problem (1384)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been returned for investigation.Root cause is unavailable at this time.
 
Event or Problem Description
Information was received that a revision procedure was performed on (b)(6) 2020.As per the reporter, the rods were corroded and would not lengthen.There was no patient injury reported.
 
Additional Manufacturer Narrative
Correction: d.4 (model number, udi number).Additional information: a.3, a.4, d.4 (lot number, expiration date), d.6, d.9, h.3, h.4, h.6, h.10.The reported failure mode was confirmed.The rod was unable to distract with manual distractor & erc.Due to the current condition of the rod (jammed), force testing was not performed.The rod was sectioned, and debris build up was found, which may have caused the reduced functionality.Sectioning of the rod determined that it could not be separated from the housing without use of high force.The cause of the jam cannot be definitively determined but may have been the bending force applied on the rods during the distraction sessions.Additionally, the patient¿s daily activities and unique anatomical structure can influence the amount of force applied to the rod.Review of the device history records revealed no discrepancies related to this complaint.The rod was manufactured in accordance with the specified requirements and met all of the required quality inspections.
 
Event or Problem Description
No additional event information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key10557801
Report Number3006179046-2020-00393
Device Sequence Number16842489
Product Code PGN
UDI-Device Identifier00812258026349
UDI-Public812258026349
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2018
Device Model NumberMS1-4590S
Device Lot NumberA160930-02-0AC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/24/2020
Supplement Date Received by Manufacturer03/25/2021
Initial Report FDA Received Date09/19/2020
Supplement Report FDA Received Date03/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Weight40
-
-