• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. PROCISE EZ COBLATOR II; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. PROCISE EZ COBLATOR II; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number EIC8870-01
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Information (3190)
Date of Event 08/14/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during an ent procedure, when the tonsil was ablated with the default gear of the "procise ez wand", the electrode fell off.After the loss, x-ray observation was made and no electrode was found in the brain.The procedure was completed with a surgical delay of 2 hours using a back-up device.No further complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found plasma wand devices are designed for ablation and/or coagulation only, and not for mechanical displacement of tissue through applied force, which may result in bent or detached electrodes or patient injury.A review of this complaint case revealed there were no patient injuries reported and no apparent patient impact based on the details provided.Therefore, no further medical assessment is warranted.The customer provided image confirms the product identification information , but provides no functional information.A relationship, if any, between the subject device and the reported event could not be determined.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCISE EZ COBLATOR II
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10558155
Report Number3006524618-2020-00798
Device Sequence Number2793701
Product Code GEI
UDI-Device Identifier00817470000450
UDI-Public00817470000450
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K070374
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/15/2022
Device Model NumberEIC8870-01
Device Catalogue NumberEIC8870-01
Device Lot Number2035319
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/27/2020
Supplement Date Received by Manufacturer10/21/2020
Initial Report FDA Received Date09/20/2020
Supplement Report FDA Received Date10/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-