• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SURESHOT HUMERAL 3.5MM HEXDRIVER; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SURESHOT HUMERAL 3.5MM HEXDRIVER; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 71691153
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/25/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to surgery, during instrument inspection was noticed that the sureshot humeral 3.5mm hexdriver is broken.The procedure was completed using a smith and nephew back up device, with no surgical delay.There was no patient involved when the issue was identified.
 
Additional Manufacturer Narrative
H3, h6: the device, intended use in treatment, was not returned for evaluation, the reported event could not be confirmed.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESHOT HUMERAL 3.5MM HEXDRIVER
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10558426
Report Number1020279-2020-04740
Device Sequence Number13767481
Product Code OLO
UDI-Device Identifier03596010656643
UDI-Public03596010656643
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K110240
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71691153
Device Catalogue Number71691153
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/25/2020
Supplement Date Received by Manufacturer12/27/2020
Initial Report FDA Received Date09/21/2020
Supplement Report FDA Received Date12/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-