Catalog Number 8065752452
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that the tip of the cutter broke and fell in the eye during surgery.The surgery was performed and completed after replacing the product with another one.The broken tip was removed from the eye within the same session.There's no patient harm.
 
Manufacturer Narrative
One opened probe was received with a tip protector, in a tray, for the report of tip broke, fell off in the eye, and was removed.The returned sample was visually inspected and found non-conforming with the tip of the probe needle broken at the port.The probe was disassembled and the components inspected.No/minimal wear was observed on the inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Gouge marks were observed at several locations along the inner cutter.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The complaint evaluation confirmed that the probe had a broken tip at the port.The exact root cause for the broken tip cannot be determined from this evaluation.An internal investigation has been completed to further evaluate the cause of probe tip breakage at the port.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints will continue to be reviewed and closely monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key10560976
MDR Text Key209599927
Report Number1644019-2020-00503
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Catalogue Number8065752452
Device Lot Number2332673H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2020
Date Manufacturer Received12/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION LXT (B)(6)
-
-