| Model Number |
R SERIES |
| Medical Device Problem Code |
Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device's pacer rate was not adjustable.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the reported customer's report was attributed to a system interconnection flex cable that was not properly seated.The cause of the loose connection appears to be from impact sustained by the device from being dropped.The damages were found to be consistent with mishandling.The cable was reseated to correct the reported problem.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|