RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Increase in Pressure (1491); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 21sep2020.
|
|
Event Description
|
It was reported that the ventilator had an error code indicating pressure regulation high.There was no patient involvement.The customer troubleshot with technical support and was advised to perform a performance verification testing (pvt).
|
|
Manufacturer Narrative
|
G4: 04nov2020.B4: (b)(6) 2020.The customer reported that a flow sensor and data acquisition board were ordered and were received.Multiple attempts were made to obtain information on the repair/service of the device that have been unsuccessful.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
G4: 04nov2020 b4: (b)(6) 2020 upon a follow up, the customer had reported to technical support that a flow sensor and data acquisition board were ordered and were received.Multiple attempts were made to obtain information on the repair/service of the device that have been unsuccessful.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|