Model Number 9735602
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/25/2020
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 9 733467, serial/lot #: n/a, version: 2. 3, patient information will not be provided due to (b)(6) confidentiality laws. No product analysis or system checkout has been performed. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation being used pre-operatively for a functional endoscopic sinus surgery (fess) procedure. It was reported that the power was turned on with all cables connected, but communication failure between the monitor and magnetic field control box occurred. The issue was corresponded by rebooting. There was a delay of less than an hour. There was no impact to patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION COMPACT
Type of DeviceEAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
tricha miles
7000 central avenue ne rcw215
minneapolis, MN 55432
7635140379
MDR Report Key10561305
MDR Text Key208595424
Report Number1723170-2020-02489
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9735602
Device Catalogue Number9735602
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/08/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-