SMITH & NEPHEW, INC. REF LNR 28X50-52 20 DEG SZ E; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
71742850 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Defective Device (2588)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Osteopenia/ Osteoporosis (2651); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/26/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a right thr procedure, after implanting the out cup 71336450 (ref three hole shell 50mm), liner 71742850 (ref lnr 28x50-52 20 deg sz e) couldn't be implanted.Try to use the liner 71333324 (ref xlpe 28 20 deg 50-52 e), still didn't work.Try to use the out cup of 71336452 (ref three hole shell 52mm), still did not work.Due to osteoporosis of the patient, the outer cup was loose and could not be tightened many times, resulting in partial loss of the posterior wall bone.Finally, the reinforce ring and bone cement cup were used to complete the surgery.In addition, there were the bone graft and hercules bone cement to prevent infection.A delay greater than 30 min was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that during a right thr procedure, the surgeon had trouble getting the cup and liner to tighten, trying with two different liners and shells.Due to osteoporosis of the patient, the outer cup was loose and could not be tightened many times, resulting in partial loss of the posterior wall bone.Finally, a reinforce ring and bone cement cup were used to complete the surgery.The affected liner, intended for use in treatment, was returned and evaluated.A visual inspection of the device shows damage on the surface of the device, probably due to attempted implantation.Our quality team completed dimensional analysis on the returned product for all critical-to-quality features.Some features were noted oversized which relate to mating features on the exterior of the part.The surface of the diameter is visually damaged around the part due to the attempted implantation.No other deviations were noted.Plastic is easily distorted during attempted implantation, especially when attempting to mate with a rigid surface like the titanium on reflection shells.Based on this information, a dimensional analysis of this part is inconclusive at this time.Our investigation including a review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.There is no information that would suggest the device failed to meet specifications.A review of the complaint history on the listed part revealed no prior complaints for the listed failure mode with the same batch number.Based on this investigation, the need for corrective action is not indicated.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.A medical analysis noted x-rays and an abbreviated surgery record were provided for review.The medical information provided about the patient was limited to her rheumatoid arthritis, severe osteoporosis, fracture of the femoral neck and several rib fractures.There was no mention of a traumatic accident or any precipitating event to cause these fractures.The impact to the patient would be the pain of the fractures, the surgery and the recovery.Based on the available information, the issues with the shells and liners was likely as the report stated, due to her severe osteoporosis.Her poor bone quality required bone grafting and cement to reinforce the bone and stabilize the components.Failure of the implants cannot be concluded.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|