RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 21sep2020.
|
|
Event Description
|
The customer contacted technical support (ts) stating that unit failed electrical safety during preventative maintenance (pm).The customer reported there was no patient involvement at the time the issue was discovered.The customer evaluated the unit and confirmed the issue to be with the unit's power cord.The customer replaced the unit's power cord to resolve the reported issue.
|
|