SMITH & NEPHEW, INC. LONG HEXDRIVER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631070
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 08/26/2020
Event Type  malfunction  
Event Description
It was reported that during inspection process, the threads of the long hexdriver were seen to be broken.There was no case involved.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirms the threads are damaged on the device.Only the hexdriver shaft was returned for evaluation.The device shows significant signs of wear/usage.This instrument was manufactured in 2015.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instruments that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG HEXDRIVER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10566426
MDR Text Key207918009
Report Number1020279-2020-04779
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010519269
UDI-Public03596010519269
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number71631070
Device Catalogue Number71631070
Device Lot Number15FM17295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2020
Initial Date Manufacturer Received 08/26/2020
Initial Date FDA Received09/22/2020
Supplement Dates Manufacturer Received09/23/2020
Supplement Dates FDA Received09/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-