MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH PRIME ROD HOLDER PUSHROD; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH PRIME ROD HOLDER PUSHROD; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 286750063
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in the (b)(6) as follows: it was reported that on (b)(6) 2020, when the surgeon finished placing the rod and tried to remove the rod holder, the rod holder would not disengage from the rod.There was a surgical delay of five (5) minutes.The procedure was successfully completed.There was no patient consequence.Concomitant device reported: setscrews (part number unknown, lot unknown, quantity unknown), screws (part number unknown, lot unknown, quantity unknown), viper prime rod holder stem (part number 286750061, lot mi63777, quantity 1), viper prime rod holder shaft (part number 286750062, lot gm5029302, lot 1), rods (part number unknown, lot unknown, quantity 1).This report involves one (1) viper prime rod holder pushrod.This is report 2 of 4 for (b)(4).
 
Event or Problem Description
Concomitant device reported: unknown setscrews (part# unknown, lot# unknown, quantity unknown).Unknown screws (part# unknown, lot# unknown, quantity unknown).This complaint involves six (6) devices.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6- the prime rod holder pushrod ((b)(6) & lot no: gm5009001) was received at us cq.Upon visual inspection, no damages were observed on the device except minimal wear which would not contribute to the complaint condition.The complaint was not confirmed for the returned device as no damages were observed on the device.The complaint device was able to be assembled with the returned mating device(s) as intended.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot a review of the receiving inspection (ri) for prime rod holder pushrod was conducted identifying that lot number gm5009001 was released in two (2) batches: batch1: lot units were released on 01-mar-2018 with no discrepancies.Batch 2: lot units were released on 25-apr-2018 with no discrepancies.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: updated data-h4 device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIME ROD HOLDER PUSHROD
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10566753
Report Number1526439-2020-01765
Device Sequence Number3578038
Product Code LXH
UDI-Device Identifier10705034507699
UDI-Public(01)10705034507699
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number286750063
Device Catalogue Number286750063
Device Lot NumberGM5009001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2020
Initial Date Received by Manufacturer 08/28/2020
Supplement Date Received by Manufacturer10/06/2020
10/21/2020
10/30/2020
Initial Report FDA Received Date09/22/2020
Supplement Report FDA Received Date10/12/2020
10/27/2020
10/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN LOCKING/SET SCREWS; UNKNOWN RODS; UNKNOWN SCREWS; VIPER PRIME ROD HOLDER SHAFT; VIPER PRIME ROD HOLDER STEM; UNKNOWN LOCKING/SET SCREWS; UNKNOWN RODS; UNKNOWN SCREWS; VIPER PRIME ROD HOLDER SHAFT; VIPER PRIME ROD HOLDER STEM
-
-