Catalog Number H938741 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/29/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a 3000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was defective.The defect was further described as a leak coming from an unknown location.This issue was identified during setup and preparation prior to patient use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added to d10, h3 and h6 (replace 4114 with 10, 3221 with 114).H10: the device was received for evaluation.Unaided visual inspection was performed which observed a small tear at the middle right side of the bag.Functional testing was performed which revealed a leak on the affected location.Additional magnified inspection was performed which verified a tear/hole where the leak was located.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|