| Model Number |
7010 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was located in the left main artery.A 3.0mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, during procedure it was noted that the catheter was fractured.The device was simply pulled out and the procedure was completed with another of the same device.There were no patient complications nor injuries reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
(e1) - initial reporter facility name: (b)(6)hospital.Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 44.6cm distal of the strain relief.There was blood in the guidewire lumen.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was located in the left main artery.A 3.0mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, during procedure it was noted that the catheter was fractured.The device was simply pulled out and the procedure was completed with another of the same device.There were no patient complications nor injuries reported and the patient's status was stable.
|
| |
|
Search Alerts/Recalls
|