| Model Number |
G48732 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report and determined ta pinhole in the balloon.A functional test was performed on the returned device.A cook dilation syringe (ds-60cc-s) was filled with water and attached to the balloon inflation port.The syringe was placed into an inflation handle, and negative pressure was applied to the balloon.After applying negative pressure, the balloon was inflated with water and the balloon would not hold pressure.A leak was observed from a pinhole near the distal end of the balloon material.There were no kinks, or any other damage to the catheter.A product-specific discrepancy that could have caused, or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.A possible contributing factor to a leakage in the balloon material is if the balloon comes in contact with a sharp object or burr in the endoscope accessory channel.Prior to distribution, all hercules 3 stage balloons esophageal are subjected to a visual inspection and functional testing to ensure device integrity.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During an esophagogastroduodenoscopy (egd), the physician used a cook hercules 3 stage balloon esophageal.The balloon would not inflate at all.A second balloon was placed down the scope, and the facility was not confident it would inflate.A third balloon was opened, and used to complete the procedure successfully.There was no reportable information at this time.The device was received for evaluation on 02-sep-2020, and a leak was observed from a pinhole near the middle of the balloon material.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|
|
|