The device, which was used in the procedure, was returned for evaluation.There was a relationship between the reported event and the device.A visual inspection was performed and showed the scope to have distal tip and fiber damage and a cracked and scratched distal lens.The root cause is contact with another source.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions a device labeling/ifu review was performed and indicated instructions for use includes recommendations, instructions and precautionary statements for proper use of the product.The risk level based on severity and final occurrence is acceptable.This failure is confirmed not originate from manufacturing, packaging, or labeling defects.No manufacturing related defects were observed.
|
|
It was reported that during knee scope procedure, the scope was scratched.A backup device was available to complete the procedure with no delay or patient injuries.Results of investigation have concluded that this unit has a cracked distal lens which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|