• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K ADVANTAGE IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1500-04
Device Problem Electro-Static Discharge (2149)
Patient Problem Insufficient Information (4580)
Event Date 09/04/2020
Event Type  malfunction  
Event Description
The recipient reportedly experienced a possible electrostatic discharge.The recipient previously experienced pain and discomfort and refused to wear the device after playing on the trampoline in (b)(6) 2019.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.Significant bone growth was noted around the implant during explant surgery.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some electrical tests from being performed.The device passed the electrical and mechanical tests performed.The failure of this device is attributed to a short from the power node to the case ground and a short from the voltage ground to the case ground at the analog chip.It is believed that electrostatic discharge (esd) led to an overload voltage, damaging structures on the case ground node inside the analog integrated circuit.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10570103
MDR Text Key208520710
Report Number3006556115-2020-01009
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016828638
UDI-Public(01)07630016828638(11)150917(17)180731
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/31/2018
Device Model NumberCI-1500-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1