Model Number 71675501
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 08/29/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to pseud arthrosis and a fracture of the device.The doctor says s+n devices were not defective.The outcome of the patient is unknown.No additional information was provided at this time.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that, per subsequent email, no relevant clinical information will be provided for inclusion in this investigation.Therefore, a thorough medical investigation of the reported pseudarthrosis, the fracture or the reported revision cannot be rendered.According to the surgeon, ¿the s&n devices were not defective.¿ per communications, the outcome of the patient is unknown and no additional information is available at this time.Should additional medical information be provided, this complaint will be re-asssessed.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury, abnormal loading of the limb or excessive forces applied to the implant.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 10S 10MM X 26CM 125D LEFT
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10570393
Report Number1020279-2020-04806
Device Sequence Number2618166
Product Code JDS
UDI-Device Identifier00885556343746
UDI-Public00885556343746
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K040212
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71675501
Device Catalogue Number71675501
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/29/2020
Supplement Date Received by Manufacturer08/05/2021
Initial Report FDA Received Date09/22/2020
Supplement Report FDA Received Date08/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age50 YR
-
-